Optical resolution of portable endoscopy under different temperature and humidity

Changzheng Li1,(), Jing Xie1, Dan Zhang1, Wenwei Hu1, Xue Li1, Ningbo Hao1, Minghao Liu1   

  1. 1. Department of Gastroenterology, Chinese People′s Liberation Army Rocket ForceCharacteristic Medical Center, Beijing 100088, China
  • Received:2020-04-07 Online:2020-08-15 Published:2020-08-15
  • Contact: Changzheng Li
  • About author:
    Corresponding author: Li Changzheng, Email:

Abstract:

Objective

To explore the effect of temperature and humidity on optical resolution of portable endoscopy.

Methods

Measure the maximum resolution of line pairs per millimeter [r(d)] of the portable endoscopy with resolution card fulfilling national standard JB/T 9328-1999. Evaluate the r(d) under temperature -10、-5、0、5、10、15、20、25、30、35、40、45、50℃ and relative humidity 10、20、30、40、50、60、70、80、90%.

Results

There was a tendency of decrease of r(d) at temperature under 0℃ or above 45℃ and also at humidity above 80%. The decrease was more obvious at cold moist environment, and less obvious at dry cold or dry hot environment. The endoscope maintains normal r(d) at temperature from 5 to 40℃ and relative humidity from 10% to 60%.

Conclusion

The portable endoscope has favorable resolution under a wide range of temperature and humidity.

Key words: Portable endoscopy, Optical resolution, Temperature, Humidity, Harsh environment